Abstract

In this paper a CMOS Instrumentation Amplifier (IA) for Electrocardiogram (ECG) acquisition systems is presented. This IA, configured as Differential Difference Amplifier (DDA), achieves high Common Mode Rejection Ratio (CMRR), high input impedance and low dc output voltage which are essential attributes for this building block. Besides, an innovative technique in CMOS analog design named Floating Gate (FG) is applied in the DDA designed to reduce offset. The presented IA is implemented in standard 0.5µm CMOS process working with α1.5V power supply. The dc offset voltage is reduced to 39µV, differential gain obtained is 39dB, CMRR reaches 136dB@10Hz and the input referred noise obtained is 3.7µV/√Hz.