Abstract

Samples were injected onto the HPLC system via either a Valco C14W.1 internal sample injector or a Rheodyne 7125 injector, both with dissimilar Vespel rotor seals, and linked to a quadrupole time-of-flight (QTOF) mass spectrometer. Alternatively, samples were infused directly into the mass spectrometer via a syringe-driver. Electrospray ionisation in positive mode was used in both cases. Experiments demonstrated significant differential adsorption of the analyte (RAC inhibitor, NSC23766) and its tri-deuterated internal standard onto the Vespel rotor seal, with the latter seeming to be preferentially adsorbed. The deuterated analogue also showed lower electrospray sensitivity, as demonstrated by syringe infusion of a mixture of the compounds, compared with their separate infusion at the same concentration. This study has demonstrated the problem of differential adsorption onto HPLC Vespel rotor seals, and that the use of a stable-isotope-labelled analogue as internal standard does not guarantee the constancy of the analyte/internal response ratio in quantitative methods. The partial solution was to work at much higher concentration where adsorption, whilst still apparent, was relatively insignificant.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.