Abstract

The effect of sample collection, storage conditions (time and temperature), and freeze-thaw cycles on the stability of free prostate-specific antigen (fPSA), PSA complexed with alpha1-antichymotrypsin (ACT-PSA), and total PSA (tPSA) in serum was studied. The analytes were quantified using immunoassays for tPSA and fPSA on the Elecsys system 2010 and a research assay for ACT-PSA on the ES system (Roche Diagnostics). The stability of the analytes was calculated as percentages of the values measured in samples 1 h after blood collection. When the samples were stored at 37 degrees C, at room temperature or at 4 degrees C, the stability of ACT-PSA was less impaired than that of fPSA. To avoid erroneous results in the determination of PSA isoforms and their corresponding ratios, serum samples should be preserved at 4 degrees C when the analysis is performed within 8 h after blood collection, or they should be stored at -80 degrees C if the analysis is not feasible during that period.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.