ObjectiveTo evaluate the biomechanical performances of the sacroiliac screw fixation of the first sacral vertebra with different lengths of screws using the Finite-Element Method.MethodsFirst, pelvic CT images were generated from a healthy volunteer, and multislice sagittal views were produced to determine the axis for the first sacral vertebra geometric osseous fixation pathway (GOFP). Subsequently, according to the geometric size and mechanical parameters of the iliosacral screw, the screw models with the same diameter of 7.3 mm and different lengths of 80 mm, 90 mm, 100 mm, 110 mm, 120 mm, 130 mm and 140 mm were built. Then the seven screws were assembled with the pelvic model. The maximum von Mises stress and the shape variables were evaluated for the pelvis and the screws.ResultsResults are shown for the pelvic and GOFP screw, respectively. The simulation results show that the maximum von Mises stress in the cortex of the pelvic ring of the pelvis with the 130-mm length screw is the lowest among the pelvic models with different screws. Moreover, the peak displacement of the pelvis with the 130-mm length screw is the smallest. These results indicate that under the standing condition, a 130-mm length screw can decrease the stress concentration and result in a more effective transfer of stress within the reconstructed pelvis. In addition, the displacement of the screw with a 130-mm length is the smallest among all the considered screws. The peak von Mises stresses in the 130-mm length screw and the cortex are still within a low and elastic range below the yielding strengths of the materials.ConclusionThrough the finite element analysis, the GOFP can be used as a safe and effective way for iliosacral screw fixation. The optimal length of the screw may be 130 mm length.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call