Abstract

BackgroundPatellofemoral joint replacement is a successful treatment option for isolated patellofemoral osteoarthritis. However, results of later conversion to total knee replacement may be compromised by periprosthetic bone loss. Previous clinical studies have demonstrated a decrease in distal femoral bone mineral density after patellofemoral joint replacement. It is unclear whether this is due to periprosthetic stress shielding. The main objective of the current study was to evaluate the stress shielding effect of prosthetic replacement with 2 different patellofemoral prosthetic designs and with a total knee prosthesis.MethodsWe developed a finite element model of an intact patellofemoral joint, and finite element models of patellofemoral joint replacement with a Journey PFJ prosthesis, a Richards II prosthesis, and a Genesis II total knee prosthesis. For each of these 4 finite element models, the average Von Mises stress in 2 clinically relevant regions of interest were evaluated during a simulated squatting movement until 120 degrees of flexion.ResultsDuring deep knee flexion, in the anterior region of interest, the average Von Mises stress with the Journey PFJ design was comparable to the physiological knee, while reduced by almost 25% for both the Richards II design and the Genesis II total knee joint replacement design. The average Von Mises stress in the supracondylar region of interest was similar for both patellofemoral prosthetic designs and the physiological model, with slightly lower stress for the Genesis II design.ConclusionsPatellofemoral joint replacement results in periprosthetic stress-shielding, although to a smaller degree than in total knee replacement. Specific patellofemoral prosthetic design properties may result in differences in femoral stress shielding.

Highlights

  • Patellofemoral joint replacement is a successful treatment option for isolated patellofemoral osteoarthritis

  • Mean stresses in the proximal regions of interest (ROI) were similar for both patellofemoral joint replacement designs and the physiological model, and slightly lower for the Genesis II total knee prosthetic design

  • Between 80° and 120° flexion, anterior ROI bone stresses for the Journey PFJ design were comparable to the physiological knee, while reduced by almost 25% for both the Richards II patellofemoral joint and Genesis II total knee joint replacement designs

Read more

Summary

Introduction

Patellofemoral joint replacement is a successful treatment option for isolated patellofemoral osteoarthritis. Previous clinical studies have demonstrated a decrease in distal femoral bone mineral density after patellofemoral joint replacement. It is unclear whether this is due to periprosthetic stress shielding. Measurements of the periprosthetic bone mineral density (BMD) using dual-energy x-ray absorptiometry (DXA) demonstrated a 15% decrease in BMD behind the anterior flange of the femoral component during the first year after Richards II patellofemoral joint replacement van Jonbergen et al Journal of Orthopaedic Surgery and Research 2012, 7:28 http://www.josr-online.com/content/7/1/28. Some investigators have used numerical models to calculate the stress distribution within the patellar components after patellofemoral joint replacement [9,10] None of these models analyzed a loaded distal femur during a squat to investigate the effect of the femoral component on the stress distribution in the periprosthetic bone

Objectives
Methods
Results
Discussion
Conclusion