Abstract

Xenon-enhanced computed tomography (Xe-CT) has been used to measure regional ventilation by determining the wash-in (WI) and wash-out (WO) rates of stable Xe. We tested the common assumption that WI and WO rates are equal by measuring WO-WI in different anatomic lung regions of six anesthetized, supine sheep scanned using multi-detector-row computed tomography (MDCT). We further investigated the effect of tidal volume, image gating (end-expiratory EE versus end-inspiratory EI), local perfusion, and inspired Xe concentration on this phenomenon. WO time constant was greater than WI in all lung regions, with the greatest differences observed in dependent base regions. WO-WI time constant difference was greater during EE imaging, smaller tidal volumes, and with higher Xe concentrations. Regional perfusion did not correlate with WI-WO. We conclude that Xe-WI rate can be significantly different from the WO rate, and the data suggest that this effect may be due to a combination of anatomic and fluid mechanical factors such as Rayleigh-Taylor instabilities set up at interfaces between two gases of different densities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.