Abstract

The objective of the study is to investigate the difference in 1-year late lumen loss (LLL) between the high- (IN.PACT Admiral) and low-dose (Lutonix) paclitaxel-coated balloon (PCB). Although a recent randomized clinical trial demonstrated no difference in efficacy endpoint between high- and low-dose PCB, it remains unclear whether high-dose PCB was superior to low-dose PCB in actual clinical practice. We enrolled 64 patients with 67 de novo femoropopliteal lesions who underwent PCB angioplasty at Kokura Memorial Hospital from May 2014 to March 2020 and subsequent follow-up angiography after 1year. The primary endpoint was 1-year LLL, whereas the secondary endpoints were binary restenosis and clinically driven target lesion revascularization (CD-TLR) after 1year. The high- and low-dose PCB groups had 45 and 22 lesions, respectively. Although the low-dose PCB group had higher rates of coronary artery disease, hemodialysis, and chronic limb-threatening ischemia than the high-dose PCB group, the latter had a longer lesion length and more lesions with a TASC classification C or D than the former. The high-dose PCB group had a significantly lower LLL than the low-dose PCB group (0.40 ± 1.05 vs. 1.19 ± 1.03mm; P = 0.003, respectively). Moreover, the high-dose PCB group had significantly lower rates of binary restenosis at 1year than the low-dose PCB group (22.2% vs. 50.0%; P = 0.02, respectively). Moreover, negative LLL was only observed in the high-dose PCB group (33.3% vs. 0%, P = 0.005). The high-dose PCB group had a significantly lower LLL than the low-dose PCB group.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call