Abstract

The present work describes the fully integration in line of a photo-degradation unit, comprising a low pressure UV lamp, in a Multipumping Flow System (MPFS), for the fluorimetric chemical control of commercially available pharmaceutical formulations containing diazepam. The utilization of an organized micellar medium provided enhanced fluorescence emission. The results allowed to obtain a linear working range for diazepam concentrations of up to 40 mg L(-1) (r = 0.9998) and the detection limit was about 0.97 mg L(-1). The results obtained by the miniaturized and automatic flow system were in agreement with those furnished by the reference procedure, with relative deviations comprised between -2.09% and 2.13%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.