Abstract
A soft X-ray beam-position monitor based on a thin CVD diamond photodetector has been developed for use on third-generation synchrotron radiation facilities, and has been tested on beamlines ID26 and ID12 at the European Synchrotron Radiation Facility. Since diamond exhibits a low absorption to low-energy X-ray photons, this device allows beam-position monitoring with very little beam attenuation at photon energies as low as 2 keV. The devices presented here find their use essentially on low-energy monochromatic beamlines with narrow beam sizes (typically below 1 mm2). Measurements performed on a 20 µm-thick device show that a position sensitivity of 2 µm can readily be achieved with a photon flux attenuation of 22% at 4 keV. This allows permanent in-situ monitoring of beam instabilities during experiments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.