Abstract

The performance of a single crystal diamond (4.7x4.7x0.5mm3 active volume) detector was tested in the ISIS pulsed neutron beam using biparametric (time of flight and pulse height) data acquisition. Three characteristic regions in the biparametric spectra are observed: i) low pulse height events with very short time of flight induced by γ-rays; ii) low pulse height events at longer flight times (i.e. neutron energies En > 3.5–6 MeV), possibly due to neutron elastic scattering off 12C; iii) events with large pulse height and flight times corresponding to En > 6 MeV mainly due to inelastic reactions such as 12C(n,α)9Be and 12C(n,n')3α. The potential use of this detector is discussed in relation to the ChipIr neutron beam line for fast neutron irradiation of electronic components at the ISIS spallation source.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.