A photon beam-position monitor using a diamond foil as a position-sensitive device has been developed for use on insertion-device beamlines of third-generation synchrotron radiation facilities such as SPring-8, and was tested on the undulator beamline of the Tristan Super Light Facility at KEK. The beam test results show that the diamond monitor can be operated in a photoconductive mode like a semiconductor detector. It has a linear working range of about +/-1 mm and a position sensitivity of less than 3 mum. The stability of the monitor was confirmed by continuous operation under low photon beam intensity conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.