Abstract

We evaluated the dialyser reprocessing agent Renalin in a 6-month prospective study of 2,759 dialyses on 59 patients. Dialysers were withdrawn after a maximum of 6 uses, and the average number of uses achieved was 4.5 Dialyser survival varied with the type of dialyser but was unaffected by the dialysate base. In vivo clearances of urea, creatinine and phosphate were not altered by reuse, but there was a small decrease in ultrafiltration characteristics of used dialysers. No clinically significant adverse event was attributed to reuse. Blood pressure was better preserved with used dialysers, and patients experienced significantly fewer intradialytic symptoms. Savings of over +A 25,000 were achieved during the study.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.