Abstract

For the most advanced multi-frame camera in the world, the shortest response time between two frames is no shorter than 1.5 ns. Thus, there is no effective way to diagnose the ultrafast sub-nanosecond dynamic of a microwave-driven plasma discharge in a single pulse. Different-length multi-sub-beam optical fibers, together with a spectrometer and an EMICCD camera, are proposed and designed to detect the nanosecond discharge spectra in a single pulse, like a real-time multi-frame spectral camera. This novel method could realize a time interval between two consecutive frames shorter than 0.1 ns by a length difference of 2 cm for sub-fibers, achieving the measurement of ultrafast plasma dynamics. Temporal evolution of electron density as well as energy of electrons and ions during nanosecond microwave discharge is further studied by de-convolving the Stark broadening and thermal Doppler broadening and by calculating the ratio of emission coefficients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.