Abstract

A system and method of processing received signals in an ultrasound imaging system is disclosed. In the system, the received signals are amplified by means of a gain controlled amplifier, and the gain of the amplifier is varied, by varying the load capacitance, as a function of depth (time) to compensate for the attenuation of ultrasonic energy at different depths within a patient's body.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.