Abstract
This study aimed to evaluate the performance of the point-of-care circulating cathodic antigen (POC-CCA) test in a highly endemic area in Brazil, comparing it to the Kato-Katz (KK) technique for sensitivity, specificity and the intensity of the reaction of the test in relation to the parasitic load. The community in Sergipe, Brazil, participated in the study, providing three stool samples, one of urine (POC-CCA) and fingers tick blood sample was tested by enzyme-linked immunosorbent assay (ELISA). Sensitivity, specificity, positive predictive value, negative predictive value, accuracy, kappa coefficient and Spearman's correlation were calculated for the POC-CCA test using the KK as the reference. The prevalence of schistosomiasis by KK testing was 48.82%; POC-CCA (t+) 66.14%; POC-CCA (t-) 45.24%. ELISA results showed 100% agreement in individuals with high and moderate eggs per gram (EPG). POC-CCA presented good diagnostic performance in individuals with medium and high EPG, but there were a high number of false negatives in individuals with low intensity infections. As observed, POC-CCA-filter test improves accuracy and sensitivity compared to a conventional test.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.