Abstract

This study aimed to evaluate the performance of the point-of-care circulating cathodic antigen (POC-CCA) test in a highly endemic area in Brazil, comparing it to the Kato-Katz (KK) technique for sensitivity, specificity and the intensity of the reaction of the test in relation to the parasitic load. The community in Sergipe, Brazil, participated in the study, providing three stool samples, one of urine (POC-CCA) and fingers tick blood sample was tested by enzyme-linked immunosorbent assay (ELISA). Sensitivity, specificity, positive predictive value, negative predictive value, accuracy, kappa coefficient and Spearman's correlation were calculated for the POC-CCA test using the KK as the reference. The prevalence of schistosomiasis by KK testing was 48.82%; POC-CCA (t+) 66.14%; POC-CCA (t-) 45.24%. ELISA results showed 100% agreement in individuals with high and moderate eggs per gram (EPG). POC-CCA presented good diagnostic performance in individuals with medium and high EPG, but there were a high number of false negatives in individuals with low intensity infections. As observed, POC-CCA-filter test improves accuracy and sensitivity compared to a conventional test.