Abstract

Atrial fibrillation is the most common arrhythmia of clinical significance and hypertension is one of its major risk factors. This study aimed to evaluate the diagnostic accuracy of an automated oscillometric blood pressure (BP) monitor with the function of atrial fibrillation detection for atrial fibrillation screening. Patients attending outpatient cardiology clinics were recruited for atrial fibrillation screening by the BP monitor with triplicate BP measurements for atrial fibrillation detection. Furthermore, a single-lead ECG was recorded simultaneously for comparison as the reference standard. The diagnostic test's evaluation index were analyzed, including sensitivity, specificity, and receiver operator characteristic (ROC) analysis. A total of 295 participants were analyzed including 166 males and 129 females, with an average age of 72.5 ± 5.9 years. The sensitivity and specificity for atrial fibrillation detection by the device were 1.000 and 0.904, respectively, with the area under the ROC curve of 0.952 (95% confidence interval: 0.929-0.975, P < 0.001). Furthermore, the device had a Kappa-value of 0.781 ( P < 0.001) with the single-lead ECG in detecting atrial fibrillation. The automated oscillometric BP monitor (G.LAB MD41A0) with atrial fibrillation detection function has high sensitivity and specificity with good accuracy for atrial fibrillation screening, which could be used as a reliable screening tool for the early detection of atrial fibrillation with potential benefits.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call