Abstract

Tablet-based perimetry could be used to test for glaucomatous visual field defects in settings without easy access to perimeters, although few studies have assessed diagnostic accuracy of tablet-based tests. The goal of this study was to determine the diagnostic accuracy of iPad perimetry using the visualFields Easy application. This was a prospective, cross-sectional study of patients undergoing their first Humphrey Field Analyser (HFA) visualfield test at a glaucoma clinic in India. Participants underwent 24-2 SITA Standard HFA testing and iPad-based perimetry with the visualFields Easy application. Reference standards for both visualfield loss and suspected glaucoma were determined by ophthalmologist review of HFA results and optic disc photographs. Receiver operating characteristic curves were constructed to assess diagnostic accuracy at various test thresholds. 203 eyes from 115 participants were included, with 82 eyes classified as moderate or worse glaucoma. iPad perimetry had an area under the receiver operating characteristic (AUROC) curve of 0.64 (95% CI 0.57 to 0.71) for detection of any visualfield defect relative to HFA and an AUROC of 0.68 (0.59 to 0.76) for detection of moderate or worse glaucoma relative to ophthalmologist examination. At a set specificity of 90%, the sensitivity of iPad perimetry for detection of moderate or worse glaucoma was 35% (22-48%). iPad perimetry using the visualFields Easy application had inadequate diagnostic accuracy to be used as a screening tool for glaucoma in this South Indian population.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call