Abstract
this study is aimed at evaluating the effectiveness of digital images for telecytology diagnosis and compares it with routine cytology diagnostic under the conditions of Georgia. gynecological cytology cases (n = 420) were taken from the clinical laboratory. Cases were diagnosed routinely by one of four certified cytologists who provided cytology diagnoses. Digital images were obtained on all cases and were evaluated as computer images by a panel of cytologists. there was 94% concordance in average between routine versus digital images diagnostic. Intracytologists concordance averaged 95.5%. Image sharpness and quality were rated "good" and "excellent" in 97% cases. With respect to image color, 96% of the images were rated as "excellent" or "good." digital images for cytology diagnostic are of adequate quality, with diagnostic concordance rates.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.