Abstract

Samples containing microM concentrations of dopamine, (+/-)-isoproterenol, para-aminophenol and chlorogenic acid have been separated by capillary electrophoresis (CE) and detected using end-column amperometric detection based on a novel decoupling method. The present decoupling approach involves the use of an electrochemical detector chip containing an array of microband electrodes where the working and reference electrodes are positioned only 10 microm from each other. The short distance between the working and reference electrodes ensures that both electrodes are very similarly affected by the presence of the CE electric field. With this method, no shift in the detection potential was seen when the CE high voltage was applied. This eliminated the need for a reoptimization of the detection potential to compensate for the influence of the separation voltage on the detection. It is also demonstrated that catecholamines can be detected using gold microband electrodes by careful adjustment of the detection potential to avoid the formation of gold oxide. Such careful adjustments of the detection potential are straightforward using the present decoupling method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.