Abstract

To evaluate a device designed for occlusion of rectovaginal fistula in patients with acquired rectovaginal fistula. The device consisted of a disk portion, an anchoring portion, and a shaft that connected each portion. The device was framed with a nitinol wire. The disk portion was framed in a four-leaf clover configuration, and the anchoring portion had two ellipsoid arms. The disk and anchoring portions were mounted with a nylon patch and coated with silicone. The device was designed for transrectal insertion and was placed in seven women with rectovaginal fistula caused by pelvic irradiation or pelvic surgery. All devices were placed successfully, and all fistulas were completely occluded after placement of the device. No procedural complications were encountered. The follow-up period was 1-26 months. One patient died of distant metastasis 5 months after placement of the device. None of the patients reported leakage from the rectovaginal fistula during the follow-up period. The occlusion device may be promising and beneficial for use in occlusion of rectovaginal fistula that results from pelvic irradiation or pelvic surgery.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.