Abstract

The development of digital hearing aids will allow for more sophisticated signal-processing techniques than have previously been implemented in commercially available analog devices. The effects of applying noise reduction or spectral enhancement, for example, must be quantified in a reasonable and consistent manner. Methods for characterizing linear aids using pure tones or stationary broadband noise are no longer sufficient for this task. Indeed, current standard techniques fail to satisfactorily describe the most sophisticated of the currently available analog devices. Techniques for characterizing nonlinear devices must be developed that capture and quantify the abilities of these new processing schemes. This is also important so that the performance differences between alternative processing schemes are readily comparable and understandable by hearing health care specialists. A set of tests is presented that characterizes the performance of DSP aids and the results are related to standard psychoacoustic measures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.