Abstract

The possibilities of computer-assisted and automated accelerator mass spectrometry (AMS) measurements were explored. The goal of these efforts is to develop fully automated procedures for “routine” measurements at the Vienna Environmental Research Accelerator (VERA), a dedicated 3-MV Pelletron tandem AMS facility. As a new tool for automatic tuning of the ion optics we developed a multi-dimensional optimization algorithm robust to noise, which was applied for 14C and 10Be. The actual isotope ratio measurements are performed in a fully automated fashion and do not require the presence of an operator. Incoming data are evaluated online and the results can be accessed via Internet. The system was used for 14C, 10Be, 26Al and 129I measurements.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.