Abstract

Fractional flow reserve (FFR) pullbacks assess the location and magnitude of pressure drops along the coronary artery. The pullback pressure gradient (PPG) quantifies the FFR pullback curve and provides a numeric expression of focal versus diffuse coronary artery disease. This study aims (1) to validate the PPG using manual FFR pullbacks compared with motorized FFR pullbacks as a reference; and (2) to determine the intra- and interoperator reproducibility of the PPG derived from manual FFR pullbacks.Patients with stable coronary artery disease and an FFR ≤ 0.80 were included. All patients underwent FFR pullback evaluation either with a motorized device or manually, depending on the study cohort. The agreement of the PPG between repeated pullbacks was assessed using the Bland-Altman method. Overall, 116 FFR pullback maneuvers (96 manual and 20 motorized) were analyzed. There was excellent agreement between the PPG derived from manual and motorized pullbacks (mean difference -0.01 ± 0.07, 95% limits of agreement [LOA] -0.14 to 0.12). The intra- and interoperator reproducibility of PPG derived from manual pullbacks were excellent (mean difference <0.01, 95% LOA -0.11 to 0.12, and mean difference <0.01, 95% LOA -0.12 to 0.11, respectively). The duration of the pullback maneuver did not impact the reproducibility of the PPG (r = 0.12, 95% CI: -0.29 to 0.49, p = 0.567).Manual pullbacks allow for an accurate PPG calculation. The inter- and intraoperator reproducibility of PPG derived from manual pullbacks were excellent.