Abstract
Chiral discrimination of thyroxine (T4) enantiomers was performed using ultra high-performance liquid chromatography–tandem mass spectrometry (UHPLC-MS/MS) on a chiral crown ether-derived ChiroSil SCA (-) column. The different composition of mobile phases and the effect of column oven temperatures were investigated and the optimum chromatographic separation with respect to resolution and analysis time was achieved using a mixture of 60% methanol/water (v/v) with 0.1% formic acid at 40 °C having a flow rate of 1.4 mL min−1. The thermodynamic data from van’t Hoff plots of temperature experiments revealed that the enantioseparation was enthalpically favored process. The method was validated in the concentration range of 0.5–100 μg mL−1 for both enantiomers and proved to be rapid, precise, sensitive, and selective method for the enantiodiscrimination of T4 under the optimized conditions. The calibration curves of both D- and L-T4 showed an excellent linearity with coefficient of determination (R2) > 0.9997. The developed chiral method was successfully applied for a quantitative assay to check the enantiomeric purities of the six pharmaceutical formulations of levothyroxine sodium tablets and the enantiomeric impurities identified were in the range of 0.11–0.29%. This method could be applied for the determination of enantiomeric purity on pharmaceuticals and also for the monitoring of thyroid hormone levels.Graphical Abstract
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.