Abstract

Abstract Sample preparation methods for four-wipe composite dust wipe samples, used by nine different laboratories accredited under the U.S. Environmental Protection Agency (EPA) National Lead Laboratory Accreditation Program (NLLAP), were evaluated to select two methods most likely to provide the best overall performance among multiple laboratories, based on accuracy, precision and practicality of the methods. The best performing methods involved (1) hot nitric acid/hydrogen peroxide digestion and (2) nitric acid ultrasonication. In a second round of testing, two sets of composite dust wipe samples with known amounts of lead were submitted blind to each of the nine laboratories to be prepared for analysis using these two sample preparation methods. Results showed that both methods are capable of meeting all EPA NLLAP requirements for accuracy (recovery) and precision. The ultrasonication method had superior performance at lower cost.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.