Abstract

SRON is developing X-ray transition edge sensor (TES) calorimeters arrays, as a backup technology for X-IFU instrument on the ATHENA space observatory. These detectors are based on a superconducting TiAu bilayer TES with critical temperature of 100 mK on a 1 μm thick SiN membrane with Au or Au/Bi absorbers. Number of devices have been fabricated and measured using a Frequency Division Multiplexing (FDM) readout system with 1-5 MHz bias frequencies. We measured IV curves, critical temperature, thermal conductance, noise and also X-ray energy resolution at number of selected bias points. So far our best calorimeter shows 3.9 eV X-ray resolution at 6 keV. Here we present a summary of our results and the latest status of development of X-ray calorimeters at SRON.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.