Abstract

The SPARC tokamak will be equipped with a hard X-ray (HXR) monitor system capable of measuring the bremsstrahlung emission from runaway electrons with photon energies in excess of about 100keV. This diagnostic will detect the formation of runaway electron beams during plasma start-up and inform the plasma control system to terminate the discharge early to protect the machine. In this work, we present a 0D estimate of the HXR emission in SPARC during plasma start-up. Then we discuss the characterization of a prototype of the HXR monitor. The detector mounts a 1 × 1-in.2 LaBr3 inorganic scintillator coupled with a photomultiplier tube and has been tested with γ-ray sources to find its dynamic range. Finally, two possible modes of operation for spectroscopic and current mode measurements on SPARC are proposed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.