Abstract

A new microbeam facility is being constructed at the 1.7 MV Pelletron Accelerator in Jyväskylä. The facility is designed for easy upgrading and incorporates a number of innovative features. Initially, it is based on a Heidelberg doublet with a design capability of a 3 × 5 μm beamspot at PIXE intensities and later upgraded to nanobeam performance. A thermal-expansion compensated rigid frame mounted on a mechanically isolated floor section is used to support the ion optical components. A compact-post focusing electrostatic deflector is used for high linearity beam scanning. This together with a novel time-stamped data collection (TDC) allows dynamic effects in IBIC, fluorescence bleaching to be studied as well as facilitating multi-resolution image support for low-fluence imaging of cells. The target chamber is fitted with a novel low-cost large working distance optical microscope, extremely compact large solid angle photon detectors as well as conventional secondary electron, PIXE and Scanning Transmission Ion Microscopy (STIM) detectors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.