Abstract

This paper introduces the development of the special traceability measuring tool for microforce value with the test force range of 0.009807 N ~ 9.807 N (1 gf ~ 1 kgf) and the accuracy of better than ± 0.1 %, including the selection of the electromagnetic balance module, the design of the hardware structure, and the compilation of the software program. Finally, the special traceability measuring tool integrating the test measurement, automatic data collection and automatic data processing is realized, and the test verification result is given. It provides a method to solve the problem of trace the small force value of existing micro Vickers hardness tester and Knoop hardness tester.

Highlights

  • With the development of modern material surface engineering, such as Physical vapor deposition, magnetron sputtering, ion implantation, surface modification, thermal spraying, nitriding and carburizing, etc., the thickness of micro Vickers hardness and Knoop hardness test samples or the surface modified layer on the sample surface is getting thinner and thinner, and the hardness test force value is getting lower and lower, reaching 0.009807 N

  • It can be seen from table 3 that the accuracy and repeatability of the development of the special traceability measuring tool for micro force value are 0.005% and 0.003%, meeting the design and use requirements

  • This paper introduces the development of a special measuring tool for micro force traceability, which has a measuring force range of 0.009807 N ~ 9.807 N (1 gf ~ 1 kgf), and the accuracy is better than ± 0.1%

Read more

Summary

INTRODUCTION

With the development of modern material surface engineering, such as Physical vapor deposition, magnetron sputtering, ion implantation, surface modification, thermal spraying, nitriding and carburizing, etc., the thickness of micro Vickers hardness and Knoop hardness test samples or the surface modified layer on the sample surface is getting thinner and thinner, and the hardness test force value is getting lower and lower, reaching 0.009807 N. The minimum measuring range of the existing force measuring instrument is 1 N. According to the principle of one tenth use, the minimum measurement can be 0.1 N, which can not meet the existing requirements. It is very necessary and urgent to develop the special traceability measuring tool with the lower limit of 0.009807 N

PROJECT DESIGN
Selection and design of electromagnetic balance module
SUMMARY

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.