Abstract

Extremity radiation monitoring is an important tool for the assessment of occupational exposures to staff at a variety of workplaces where ionising radiation is used. This work shows the feasibility of applying 3D printing for the development of customisable ring dosemeters. The rings were developed using two types of resin, hard and flexible and has the possibility of sterilisation using different techniques. The printed ring dosemeter was associated with BeO optically stimulated dosemeters. The energy and angular response were found within ±20% in the energy range from 24 to 662keV and from 0° to 60° angle of incidence. This contributes to the reduction of measurement uncertainty when compared with currently used thermoluminescent detectors dosemeters. The new ring dosemeter showed a satisfactory response with respect to the performance criteria of the IEC 62387 Standard, in addition to providing improved ergonomics in relation to the commercial ring dosemeter.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.