Abstract

We have developed PLEMO-P3, a rehabilitation system for the upper limbs that utilizes ER brakes as torque generators and is a passive haptic device. To detect abnormalities in patients, we developed a new sensing device and installed it on the grip of the PLEMO-P3. In this paper, we compare several kinds of information from the sensor system (grasping force, reaction force against the working plane, and wrist joint range) and traditional stroke assessment (BRS, FMA and STEF). We establish effective assessment and training methods with PLEMO-P3, finding the effects of training and a strong correlation between them. We suggest a training method with PLEMO depending on the results of clinical assessment. We confirmed the correlation of the working plane reaction force and wrist joint range of motion and were able to detect synergy movement within the motion sensor wrist joint range. This correlation has been understood that it will equip it only with the wrist joint range of motion sensor in PLEMO-P4, a machine for practical use.11. This paper is the full translation from the transactions of JSME, Series C, Vol.76, No.762, pp. 323-330, 2010.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.