Abstract

We report on the development of quantitative, reproducible diagnostic observables for age-related macular degeneration (AMD) based on high speed spectral domain optical coherence tomography (SDOCT). 3D SDOCT volumetric data sets (512 x 1000 x 100 voxels) were collected (5.7 seconds acquisition time) in over 50 patients with age-related macular degeneration and geographic atrophy using a state-of-the-art SDOCT scanner. Commercial and custom software utilities were used for manual and semi-automated segmentation of photoreceptor layer thickness, total drusen volume, and geographic atrophy cross-sectional area. In a preliminary test of reproducibility in segmentation of total drusen volume and geographic atrophy surface area, inter-observer error was less than 5%. Extracted volume and surface area of AMD-related drusen and geographic atrophy, respectively, may serve as useful observables for tracking disease state that were not accessible without the rapid 3D volumetric imaging capability unique to retinal SDOCT.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.