Abstract
Our study recalculated the use factor of linear accelerators (LINACs) by using an in-house program based on Digital Imaging and Communications in Medicine radiation therapy (DICOM-RT). We considered the impact of advancements and changes in treatment trends, including modality, technology, and radiation dose, on the use factor, which is one of the shielding parameters. In accordance with the methodology described in the NCRP 151 report, we computed the use factor for four linear accelerators (LINACs) across three hospitals. We analyzed the results based on the treatment techniques and treatment sites for three-dimensional conformal radiation therapy (3D-CRT) and intensity modulated radiation therapy or volumetric modulated arc therapy. Our findings revealed that the use factors obtained at 45° and 90° were 14.8% and 13.5% higher than those of the NCRP 151 report. In treatment rooms with a high 3D-CRT ratio, the use factor at a specific angle differed by up to 14.6% relative to the NCRP 151 report value. Our results showed a large difference in the use factor for specific sites such as the breast and spine, so it is recommended that each institution recalculate the use factor using patient's data.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.