Abstract

To address the issue of the difficulty in implementing non-invasive continuous blood pressure measurement technology in China, this study developed a high-performance synchronous electrocardiogram (ECG) and photoplethysmography (PPG) signal acquisition system. A PC-based human-computer interaction software platform was constructed, and continuous blood pressure measurement-related algorithms were integrated. Multiple feature parameters such as pulse wave transit time based on synchronous ECG and PPG signals were extracted, enabling non-invasive continuous blood pressure measurement. To verify the measurement accuracy of the system, tests and comparative verifications were carried out. The results demonstrated that the system could meet the requirements of relevant standards and have good application value.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.