Reliable tracheal or tissue graft has not been developed yet for the reconstruction of large, circumferential tracheal defects. Major limitations were anastomotic dehishence and stenosis, which were attributed to the poor epithelisation of the prosthetic graft. We developed a new tracheal prosthesis that has a viable lined and well-vascularized mucosa. The prosthesis consists of Prolene mesh reinforced with polypropylene rings, and is coated with gelatin. In addition, we lined the luminal surface of the prosthesis with transplanted autogenous oral mucosa and wrapped the prosthesis with greater omentum. Animal experiments were performed using 10 adult mongrel dogs. The transplanted mucosa and wrapped greater omentum tightly adhered to the prosthesis to make a single unit within two weeks. The mucosa survived well, was well vascularised by new vessels from greater omentum and showed normal histology. Complete surgical resection and replacement of a thoracic trachea (3 cm in length, 6 tracheal rings) were carried out in 2 dogs, which survived well with normal activity. We concluded that this highly biocompatible tracheal prosthesis could be very useful for step-wise reconstruction of tracheal defects.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.