Abstract
Accurate quantitative analyses require standardized methods to control and improve the analytical process in the laboratory. The availability of urine reference materials (RMs) may offer a feasible option to improve the accuracy of urine analysis and to control matrix effects. This paper presents the complete process of the development of matrix RMs in urine, including sample preparation, homogeneity, and stability studies, as well as uncertainty assessment. A freeze-drying process was developed, and freeze-dried human and pig urine samples were prepared and verified to have comparable homogeneity to liquid samples and higher stability than liquid human, pig, and artificial urine samples at 4℃ or room temperature and under extreme conditions. A total of 21 authentic urine samples from August 2022 were measured with freeze-dried RMs and spiked urine samples, and the reliability of the quantification of the RMs was compared. The freeze-dried human urine matrix RM appeared to be an excellent tool for daily quality control, as it showed high stability and gave the most consistent results with spiked samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.