Abstract
A Si photodiode detector PIN-type 10/spl times/10 mm, with a slim film of a converter material capable of producing charged particles, was used as sensor of neutrons and it was projected to be used in an environment of a zero power reactor. Polyethylene n(CH/sub 2/), which produces recoil protons from the (n,p) interaction, was used to improve the detection efficiency. A mathematical model was applied to fit the experimental data and the optimal thickness of the converters was determined by the maximum point of the function. For an AmBe source, the optimum polyethylene layer thickness was of 0.105 cm (96.6 mg/spl middot/cm/sup -2/). The converter of polyethylene was capable to improve the detection efficiency to a factor of 16.4 compared to measurement without converter. The detector shows a good response (r=0.999372) to estimate the power variation during the operation of the zero power-type reactor (IPEN/MB-01).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.