Abstract

A novel device with microchannels for flowing cells and twin microcantilever arrays for measuring the electrical impedance of a single cell is proposed. The fabrication process is demonstrated and the twin microcantilever arrays have been successfully fabricated. In our research, we measured the electrical impedance for normal and abnormal red blood cell over the frequency range from 1 Hz to 10 MHz. From the electrical impedance experiment of normal and abnormal red blood cell, it was examined that the electrical impedance between normal and abnormal red blood cells was significantly different in magnitude and phase shift. In this paper, we show that the normal cell can be taken apart from the abnormal cell by electrical impedance measurement. Therefore, it is expected that the applicability of this technology can be used in cellular studies such as cell sorting, counting or membrane biophysical characterization.