Abstract

We propose to minimize the sampling time for high-throughput measurements of powder X-ray diffraction (XRD) and X-ray absorption fine structure (XAFS) in synchrotron radiation. The conventional synchrotron radiation powder X-ray diffraction method requires filling of a capillary tube, but a structure-refining diffraction pattern could be obtained by transferring the crushed powder to a tape and rotating the cassette-tape tool by ±5° around the sample position. XAFS spectra could also be measured with the sample attached to the tape. The time required for sample preparation was greatly reduced, which made high-throughput experiments with powders in synchrotron radiation experiments more accessible.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.