Abstract

Integrating the SQUIDs and sensing coils of magnetic microcalorimeters onto the same die is a promising approach for maximizing flux coupling and signal/noise. However, new challenges in microfabrication must be overcome, because the underlying SQUID devices are sensitive to chemical attack and elevated processing temperatures. In this report, we describe development and details of a microfabrication process for integrated SQUID/sensor gamma-ray magnetic microcalorimeters with electroformed gold absorbers, starting from a modified version of the STAR Cryoelectronics “Delta 1000” Josephson Junction process.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.