Abstract

As part of the DIII-D diagnostic complement designed to address L–H transition physics issues, we have developed and commissioned a diagnostic neutral lithium beam and multichannel fluorescence detection system to investigate the edge plasma density and its associated fluctuations. The use of lithium offers several advantages for tokamak edge beam emission spectroscopy studies, including large excitation cross sections which are relatively insensitive to temperature variation, the availability of the 670.8 nm resonance line well separated from most plasma line emission, and the suitability of modest beam energies and currents to probe even dense H-mode plasmas. These features permit measurements of collisionally induced fluctuations to be obtained with good spatial (<1 cm) and temporal (<10 μs) resolution. The improvements over previous lithium beam diagnostics which were required to successfully make these measurements in a large, remotely controlled machine environment will be discussed, along with the present state of the diagnostic system and our plans for future improvements of this technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.