Abstract

We developed and evaluated the properties of in-house urine reference materials for the verification of laboratory refractometers, which are frequently used in clinical chemistry and doping testing laboratories. Urine was gathered from 26 healthy volunteers (16 male 30±5years old and 10 female 29±4years old), from which two urine batches were obtained: one with a low specific gravity (1.012±0.003) and the other with a high specific gravity (1.027±0.003). Homogeneity studies were conducted over 20 consecutive days. For short-term stability studies, aliquots of both urine batches were stored at -20±2°C; 3±2°C; 20±2°C; 45±2°C for 0, 2, 7, 14 and 35days, under both light and dark conditions. Similarly, another study was conducted to measure the long-term stability of urine at -20±2°C, over a 24-month evaluation period. Our data showed that the urine was homogeneous and stable at -20±2°C, 3±2°C, 20±2°C, and 45±2°C under both light and dark conditions. In all cases, the urine was evaluated by specific gravity and no statistically significant differences (p≤0.05) were recorded. Additionally, a proficiency test was conducted in collaboration with 15 ISO/IEC 17025 accredited laboratories, and z-scores and performance factors were evaluated. These data indicate that this material could be used for the verification of refractometers.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call