ABSTRACTHuman serum certified reference material (CRM), NMIJ CRM 7407-a, for the analysis of polychlorinated biphenyls (PCBs) was developed by the National Metrology Institute of Japan. A pool of commercially available human serum was used as a raw material of the CRM. This sample is in the form of a liquid comprising approximately 4 g stored in a cryogenic polypropylene vial. Homogeneity assessment was performed, and the material was homogeneous enough for PCB 118, PCB 138, PCB 153, and PCB 194: the relative uncertainties due to inhomogeneity were 2.5–10.5%. The results obtained from the stability assessment indicated that the target PCBs were stable: the relative uncertainties due to instability were 0–14.7%. The certification was carried out using two different types of GC columns for each target PCB to avoid interferences on GC separation; the certified values of the target PCBs (PCB 118, PCB 138, PCB 153, and PCB 194) were 9.7–129.8 ng/kg. This is the first frozen human serum CRM in which PCBs were determined by isotope dilution mass spectrometry.