Abstract
Hematocrit (HCT) is actually a calculated value obtained from modern automated hematology analyzers. It is the product of the mean cell volume (MCV) and the red blood cell (RBC) count, both of which are directly measured by the analyzer. Therefore, if there are any inaccuracies in measurement of the MCV or RBC count, the HCT will reflect those inaccuracies. In this study we have demonstrated the feasibility of electrochemical methods for determination of hematocrit level in whole blood samples. The sensitivity of the hematocrit biosensor was 1.01μA/% and the average coefficient of variation was 5.04%, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.