Abstract

Objective: To development of ethylene glycol samples in silicone tubes, and the samples were applied to inter-laboratory comparison, through it to evaluate the detection capacity of occupational health testing laboratories. Methods: Three content levels of ethylene glycol samples in silicone tubes were prepared, and the homogeneity and stability of samples were investigated, the results were statistically analyzed. A robust statistical four-point distance method was used to calculate the results submitted by each participant laboratory and the test capability of the laboratory was assessed by the z-score method. Results: The statistic of the homogeneity of ethylene glycol samples in silicone tubes were less than the critical value (P>0.05) , and the samples were stable at room temperature for one month. The satisfactory rate of comparison between 204 laboratories was 88.24%. Conclusion: The homogeneity and stability of the ethylene glycol samples in the silicone tubes can meet the requirements of proficiency testing and can be used for laboratory comparison. Most of the laboratories that participated in the comparison had the ability to detect ethylene glycol.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.