Abstract

This paper presents an entropy-based algorithm for the detection of QRS complexes (cardiac beat) in the 12-lead Electrocardiogram (ECG) using support vector machine (SVM). Digital filtering techniques are used to remove power line interference and baseline wander in the ECG signal. Entropy criterion is used to enhance the QRS complexes. SVM is used as a classifier to delineate QRS and non-QRS regions. The performance of the proposed algorithm was tested using 12-lead real ECG recordings from the standard CSE ECG database. The numerical results indicated that the algorithm achieved 99.93% of detection rate. The percentage of false positive and false negative is 0.87% and 0.06%, respectively. The performance of the proposed algorithm is found to be better than published results of the other QRS detectors tested on the same database.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.