A method for fabricating disposable membrane hydrophones is presented. The disposable hydrophones are intended for onetime use in such damaging environments as chemically contaminating fluids and high-amplitude (peak amplitude ∼100MPa) shock wave fields, where the use of commercial membrane hydrophones is not recommended. Fabrication of a hydrophone is done using only off-the-shelf components and hand tools, which translates into ease of fabrication and orders-of-magnitude reduction in unit cost. In particular, poling and sputtering, the two processes that are chiefly responsible for the cost and difficulty associated with the conventional fabrication method, are replaced with the use of pre-poled polyvinylidene fluoride (PVDF) films and polyethylene terephthalate (PET)-coated aluminum foils, respectively. Despite the seemingly crude construction, these disposable hydrophones can exhibit voltage sensitivity response that compares favorably with that of commercial hydrophones. For example, one prototype having a 2mm×2mm active element shows the end-of-cable voltage sensitivity of −270 (±1.9) dB re 1V/μPa over the frequency range of 1–10MHz.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.