Abstract

A delay line readout system for the instrumentation of X-ray multiwire gas proportional chambers has been used successfully since 1986. Developments in analogue to digital converter and sample and hold technology have made possible significant re-engineering of the system to provide better resolution, linearity and rate. The resolution has been improved from 512 to 2048 pixels while the differential linearity has been reduced from +/−90% to +/−2%. The system is designed to work with delay lines ≥ 100 ns, giving mean data rates of up to 1.5 MHz of random X-ray events. Images can be collected in 1D or 2D and a pileup rejector is included which reliably rejects multiple or partial events.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.