AbstractQuantification of cyclophosphamide, palcitaxel and etoposide in water sample using simple analytical technique is challenging. In this study, we have developed a simple, sensitive, robust and accurate method for the detection and quantification of cyclophosphamide, etoposide, and paclitaxel using high‐performance liquid chromatography and validated the same. The analytical conditions were achieved by the high‐performance liquid chromatography system on a C18 column (250 mm × 4.6 mm × 5 µm) under isocratic mode. The retention time for cyclophosphamide, paclitaxel and etoposide were 2:40.1, 3:32.3 and 3:49.3 min respectively. We validated the method by evaluating system sensitivity and linearity. We anticipate the method to be useful in quantification of cyclophosphamide, paclitaxel and etoposide in water sample.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.