Abstract

An on-line purge and trap system for continuous monitoring of Volatile Organic Compounds (VOC) is presented. The purge chamber was designed for continuous extraction of VOC from water with nitrogen. The analytes were preconcentrated on a microtrap prior to analysis by GC with flame ionization detection. The microtrap served as a fast injection device for carrying out analysis at high frequency. Continuous monitoring was accomplished by performing injections at fixed intervals. This system showed high sensitivity, high precision, detection limits at the ppb level, and stable response over long periods of continuous operation. Factors affecting system performance were studied. A predictive model based on gas-liquid partitioning is also presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.